IM 11Y01D01-01EN 7/7
3rd Edition
(2) Installation
• The installation of the equipment should be carried out by the engineers or other
professional personnel of the related expertise. The installation should not be carried out
by operators or other unprofessional personnel.
• Installation, use and maintenance of the product should comply with the manual and
following standards.
GB 3836.13-2013 “Explosive atmospheres Part 13: Equipment repair, overhaul and
reclamation”
GB 3836.15-2000 “Electrical apparatus for explosive gas atmospheres Part 15:
Electrical installations in hazardous area (other than mines)”
GB3836.16-2006 “Electrical apparatus for explosive gas atmospheres Part 16:
Inspection and maintenance of electrical installation (other than mines)”
GB 50257-2014 “Code for construction and acceptance of electric equipment on fire and
explosion hazard electrical equipment installation engineering”
• The equipment must only be used in an area of not more than pollution degree 2, as
defined in GB/T 16935.1.
• When the cable glands and/or adapters are used, they must be Ex “d”, Ex “e”, or Ex “n”
type, and must also meet the ingress protection requirement of IP54 or more. If YH8000
is installed in where more than IP54 is required as ingress protection, the cable gland
and/or adapters should be meet the requirement.
• The thread type of the cable entries shall be confirmed according to the user’s manual.
• Screws of the field wiring terminals must be tightened with the specified torque value:
0.22 to 0.25 Nm.
• Field wiring for ethernet communication must be in accordance with IEEE 802.3 so as to
avoid overvoltage of > 119 V.
WARNING
- DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE MAY BE PRESENT
- POTENTIAL ELECTROSTATIC CHARGING HAZARD
- FOR INSTALLATION AND SAFE USE, READ IM11Y01D01-01
(3) Operation
WARNING
• Electrostatic charge may cause an explosion hazard. Avoid any actions that cause the
generation of electrostatic charge, such as rubbing with a dry cloth on coating face of
product.
• Take care not to generate mechanical sparking when access to the instrument and
peripheral devices in hazardous locations.
(4) Maintenance and Repair
WARNING
The instrument modification or parts replacement by other than authorized Representative of
Yokogawa Electric Corporation is prohibited and will void the certification.
When opening the cover, the enclosure should be dry and clean to prevent from ingress water
or dust.
Содержание TDLS8000
Страница 1: ...User s Manual TDLS8000 Tunable Diode Laser Spectrometer IM 11Y01D01 01EN IM 11Y01D01 01EN 3rd Edition...
Страница 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Страница 25: ...Blank Page...
Страница 45: ...Blank Page...
Страница 115: ...Blank Page...
Страница 151: ...Blank Page...
Страница 231: ...Blank Page...
Страница 239: ...Blank Page...
Страница 243: ...Blank Page...
Страница 261: ...Blank Page...
Страница 303: ......
Страница 305: ......
Страница 307: ......
Страница 309: ......
Страница 311: ......
Страница 313: ...Blank Page...