Installation altitude: 2000 m or less
Installation category: I (Anticipated transient overvoltage 330V)
Measuring category: O (Other)
Pollution degree:
2, Indoor/Outdoor use
Note: Installation category, called overvoltage category, specifies impulse withstand voltage. Pollution degree indicates the
degree of existence of solid, liquid, gas or other inclusions which may reduce dielectric strength.
EMC Conforming Standards:
CE
EN55011 Class A Group 1
EN61326-1 Class A Table 2 (For use in industrial location), EN61326-2-3
RCM
EN55011 Class A Group 1
KC
KN11 Class A Group 1, KN61000-6-2 (Korea Electromagnetic Conformity)
한국 전자파적합성 기준
A급 기기 (업무용 방송통신기자재)
이 기기는 업무용(A급) 전자파적합기기로서 판매자 또는
사용자는 이 점을 주의하시기 바라며, 가정외의 지역에서
사용하는 것을 목적으로 합니다.
Cable conditions:
• Power cable Use a shielded cable.
• I/O cable
Use a shielded cable.
• Cable for connecting between units
Use a separately sold dedicated cable (shielded cable). Attach the ferrite clamp
supplied with TDLS8000 to the cable connecting between units on the inside of
the equipment in the case of both an SCU and LU.
• Ethernet cable Use an STP cable (shielded) of category 5e or higher.
Influence in immunity environment (criteria A):
Variation of gas concentration measurement value
Within ±15% of reading
l
YH8000 HMI Unit
Safety Conforming Standards:
CE EN61010-1
UL UL61010-1
CSA
CAN/CSA-C22.2 No.61010-1
GB
GB30439 Part 1
Installation Altitude: 2000 m or less
Installation category: I (Anticipated transient overvoltage 330 V)
Pollution degree:
2, Indoor/Outdoor use
EMC Conformity Standards:
CE
EN55011 Class A Group 1
EN61326-1 Class A Table 2 (For use in industrial location)
RCM
EN55011 Class A Group 1
KC
KN11 Class A Group 1, KN61000-6-2 (Korea Electromagnetic Conformity)
한국 전자파적합성 기준
A급 기기 (업무용 방송통신기자재)
이 기기는 업무용(A급) 전자파적합기기로서 판매자 또는
사용자는 이 점을 주의하시기 바라며, 가정외의 지역에서
사용하는 것을 목적으로 합니다.
Cable conditions:
• Power cable Use a shielded cable.
• Local HMI connection cable
Use a separately sold dedicated cable (shield cable).
• Ethernet cable Use an STP cable (shielded) of category 5e or higher.
Product conformity assessments of YH8000 for the relevant standards are performed in its own
right.
3rd Edition: Jun. 10, 2016-00
Содержание TDLS8000
Страница 1: ...User s Manual TDLS8000 Tunable Diode Laser Spectrometer IM 11Y01D01 01EN IM 11Y01D01 01EN 3rd Edition...
Страница 10: ...ix IM 11Y01D01 01EN LT LV PL EST SLO H BG RO M CZ SK 3rd Edition Jun 10 2016 00...
Страница 25: ...Blank Page...
Страница 45: ...Blank Page...
Страница 115: ...Blank Page...
Страница 151: ...Blank Page...
Страница 231: ...Blank Page...
Страница 239: ...Blank Page...
Страница 243: ...Blank Page...
Страница 261: ...Blank Page...
Страница 303: ......
Страница 305: ......
Страница 307: ......
Страница 309: ......
Страница 311: ......
Страница 313: ...Blank Page...