40
Instructions for use - Tiva2 - Tiva2H Thermodisinfector
5.5.1 Refill procedure
For the TIVA2 thermodisinfector
:
- Pull and extract the drawer manually (A Fig. 11);
- Loosen the cap with container 1 tube (or 2/3);
- Fill the container with the related chemicals.
- «Detergent»
- «Neutraliser / Rinse aid».
Tighten the cap well and put the drawer back in its position.
For TIVA2H thermodisinfector
:
- Set up a new tank with the chemical;
- Extract the empty tank 1 or 2 or 3 (Fig. 11);
- Remove the tube and place it on a non delicate and easy to
clean surface;
- Insert the tube in the relative tank and put it back in the cor-
responding position:
- «Detergent»
- «Neutraliser»
- «Rinse aid».
- At the end, close the front door with the key.
- Start the program «Fill DOS1» for the detergent or «Fill
DOS2» or «DOS3» for the neutraliser or rinse aid.
Be careful not to reverse the position of the liquids.
Failure to follow these recommendations may damage the
thermodisinfector.
- With regard to the maximum dosage for each program, fol-
low the instructions of the chemical manufacturer.
- To ensure efficiency of the dosing system, it is necessary to
regularly carry out the maintenance operations set out in the
«Maintenance» chapter.
- The indications concerning the storage and disposal of
chemical substances are provided by the respective manu-
facturers and must be observed.
- Do not place the chemical containers on the thermodisinfec-
tor.
Completely empty the thermodisinfector before performing
maintenance and before moving the machine to avoid contact
with the chemicals and to protect the machine components.