Instructions for use - Tiva2 - Tiva2H Thermodisinfector
6.2.2 Preparing the load
WARNING. The maximum weight that the open door can
withstand is 12 Kg.
The maximum volume available is 0.065 mW or 65 L. The
maximum volume of the basket is 0.025 mW or 25 L.
The instruments to be washed must be placed on the relative
basket, making sure that they do not overlap on one another.
- Empty any residual liquids from the containers before in-
serting them into the tank, thoroughly washing any residue,
e.g. disinfectant solutions.
- Insert the individual instruments in the appropriate supports
or baskets and never directly into the base basket.
- Make sure that the instruments do not come out of the con-
tainers.
- Make sure that the instruments do not hang from the grid
bars.
- Insert the hollow instruments into the appropriate injectors.
- Insert the other instruments into the adapters with silicone
insert.
- The rinse arms must be free to rotate.
- The cleaning quality depends on the correct loading of the
instruments.
- Position container-shaped objects so that liquids can flow
easily and, if possible, place tall and heavy tools in the cen-
tre of the trolley.
- Components with special geometries must be arranged so
as to allow the water to flow out.
- Introduce only suitable steel instruments, resistant to corro-
sion.
- Introduce only plastic components thermostable up to
95°C.
- To prevent possible corrosion, we recommend using only
stainless steel instruments and tools, suitable for washing.
- Nickel-plated and chrome-plated and/or aluminium instru-
ments and tools may not always be suitable for machine
treatment.