51
Using the machine
6.2.5 Treatment of ophthalmic instruments
- Only insert instruments designed for automatic treatment in
thermodisinfector. Carefully follow the instructions provided
by the instrument manufacturer.
- Do not insert instruments intended for interventions on the
optic nerve and which come into contact with the retinal
tissue.
In case of damage or alteration of the instruments despite
compliance with the manufacturer’s instructions, the re-
sponsibility will still be borne by the instrument manufactur-
er.
WARNING. The treatment of ophthalmic instruments re-
quires the use of purified water.
Comply with the following for automatic treatment:
- Use a slightly alkaline liquid for cleaning. Use a citric ac-
id-based neutraliser for neutralisation. Never use liquids for
rinsing.
- Rinse the hollow instruments after the application and
check that the purified water passage is free before the au-
tomatic treatment.
- Insert the hollow instruments into the rinse bar, specially
designed for this use.
- Make sure that no deposits form on the instruments.
- Dry the hollow instruments with compressed air after the
treatment in order to remove all residue humidity.
- Follow the manufacturer’s instructions with regard to the
maintenance of instruments/accessories for loading.