ALARM DESCRIPTION
ACTION
E01: BLACKOUT
Power failure during the cycle
Press RESET
and the program blocked.
E02: DOOR OPEN
The door is open or unlocked. Close it.
Press RESET
If it persists, please contact a technician.
E02: DOOR OPEN
Close it.
Press RESET
If it persists, please contact
a technician.
E07: NO DOOR LOCK
The door did not lock within the time
Press RESET
preset.
If it persists, please contact
a technician.
E09: NO DOOR UNLOCK The door did not unlock within the time
Press RESET
preset.
If it persists, please contact
a technician
E11: DIRTY WATER
Check for dirt residues on the
Press RESET
load and check the purified water
If it persists, please contact
in the network.
a technician.
E20: NO COLD WATER
The cold water valve could be
Press RESET
closed or partially open. Check it.
If it persists, please contact
a technician.
E21: NO HOT WATER
The hot water valve could be
Press RESET
closed or partially open. Check it.
If it persists, please contact
a technician.
E22: NO DEMI WATER
The purified water valve could
Press RESET
be closed or partially open.
If it persists, please contact
Check it.
a technician.
E26: PRINTER PAPER
The printer ran out of paper. Insert a
Press RESET
new roll.
If it persists, please contact
a technician.
E30: FLOWMETER 1
Tank 1 of the chemical liquid could
Press RESET
be empty. Check it.
If it persists, please contact
a technician.
E31: FLOWMETER 2
Tank 2 of the chemical liquid could
Press RESET
be empty. Check it.
If it persists, please contact
a technician.
E32: FLOWMETER 3
Tank 3 of the chemical liquid could
Press RESET
be empty. Check it.
If it persists, please contact
a technician.