36
Instructions for use - Tiva2 - Tiva2H Thermodisinfector
5.4.3 Regeneration rooms
The Thermodisinfector is supplied without regeneration salt
which will feed the device automatically during the regener-
ation process.
The salt must be introduced through the tank outlet inside the
tank (4 Fig. 9). To introduce the salt, unscrew the cap (coun-
ter-clockwise) and fill the tank with salt, paying attention not to
spread salt outside the bowl then close the tank with the cap.
You must fill the salt tank every time the message “Refill salt”
is displayed.
- Use only appropriate coarse-grain salt.
- Do not use kitchen salt, crushed salt tabs or other types
different from those indicated, as it may contain insolu-
ble substances.
- Do not pour cleaning liquids or other solutions in the
tank salt.
Failure to comply with these recommendations can lead to a
malfunction of the water softening device. Just before the salt
runs out completely, the display will show the message «Lack
of salt». At this point, the salt should be topped-up as soon as
possible, otherwise an error message will appear and it will no
longer be possible to activate a new cycle if not by resetting.
Proceed as follows to refill the salt:
1) Delete the message on the display by pressing the RESET
key for 5 seconds.
2) Open the door and remove any already inserted load.
3) Loosen the salt tank cap and insert the funnel.
4) In the first filling, introduce ½ litre of water to dissolve the
residue salt. FIRST FILLING ONLY
5) Fill with salt up to the edge. The tank contains about 1 kg
of salt.
6) Thoroughly clean the edge of the tank.
7) Tighten the cap again.
8) Start the «Rinse» cycle.
⇒