32
Instructions for use - Tiva2 - Tiva2H Thermodisinfector
a few minute drain in order to eliminate any impurities, air
bubbles and/or anything else could damage the machine
and obstruct its filers.
- Connect the cold and purified (if available) water hoses, exit-
ing the machine with their respective network connections, as
shown on the installation diagram. It will be the responsibility
of the installer to make sure that the temperature of the cold
water supply is not higher than 5/15°C, otherwise proper
washing of materials cannot be ensured. The connections for
cold and purified water must not feed any equipment other
than the instruments washer. During the washing cycle, this
is necessary to prevent the subdivision of the water supply
with other users, thus leading to a substantial increase in
the time required to fill the tank (in this case an alarm will be
triggered to alert the user that the maximum time allowed for
water loading is exceeded).
- If the machine is provided with a purified water feeding sys-
tem but the plant is not equipped with it, the cold and purified
water hoses should be connected together.
- The machine can be equipped with a built-in softener, which
has the function of reducing the scale in the water supplied
- Connect the flexible hoses to the machine valves posi-
tioned in view at the back, making sure to connect them
correctly based on the sales configuration.
Make sure to connect the water flexible hoses in the posi-
tions shown in figure 8.
• Connect the cold water flexible hose to connection 1 in fig-
ure 8.
• Connect the purified water flexible hose (if any), to connec-
tion 4 in figure 8.
WARNING. The flow pipes must not be shortened or
damaged.
The water supply taps must be capable of quickly stopping the
water line, therefore, they must be