30
Instructions for use - Tiva2 - Tiva2H Thermodisinfector
Additional indications regarding the electrical connection are
shown in the installation diagram. This can also be down-
loaded from www.tuttnauer.com.
5.2.1 Electromagnetic compatibility (EMC)
The machine has been tested on electromagnetic compatibility
(EMC) pursuant to Standard EN 61326-1 and is suitable for
operation in institutes such as hospitals, medical practices,
laboratories and environments connected to the public elec-
tricity grid.
The high-frequency (HF) energy emissions of the machine are
so small that interferences with electrotechnical equipment in
the immediate vicinity are not likely.
The optimal positioning floor must be made of concrete, wood
or ceramic tiles. In case of machine operation on floors made
of synthetic materials, the relative humidity must be 30% to
minimise the likelihood of electrostatic discharges.
Fig. 8 - Connections (rear view of the machine)
1) Cold water connection. - 2) Power supply socket. - 3) Drain
connection. - 4) Cold or purified water connection (depending
on the sales configuration).
1
2
3
4