27
Installation
5. Installation
It is advisable that only furniture for professional use be posi-
tioned in the area around the machine, to avoid ruining them
due to possible leakage of condensation water.
HAZARD: Do not install and/or use the thermodisinfector
in environments with flammable/explosive atmosphere.
Make sure that the floor is fit to support the load of the
equipment when in operation, equal to 70 Kg.
Make sure that the machine is perfectly vertical and sta-
ble, using a spirit level, if necessary.
Unevenness of the machine surface and height can be adjusted
with the four adjustable feet placed under the machine itself.
There are two models of thermodisinfector (see Fig. 2):
The
TIVA2
model is the classic model of thermodisinfector with
small-sized product bowls, which can be positioned on any
piece of furniture capable of withstanding its weight, within a
dental practice.
The
TIVA2H
model is a thermodisinfector with much more capa-
ble product bowls than the TIVA2 version and which, usually, is
embedded under a piece of furniture where appropriate space
is already provided, within a dental practice.
HAZARD. All electrical and water connections (loading/
unloading) can only be carried out by specialised and
authorised personnel and by consulting the relative
diagrams.
Before machine positioning make sure that:
All the components needed for installation and proper use of
the machine were installed: main switch, water supply taps,
drain and