48
Instructions for use - Tiva2 - Tiva2H Thermodisinfector
They require special process conditions.
- Objects made completely or partially of plastic must be re-
sistant to high temperatures.
- Before loading the machine, remove substances such as
composites, cement and amalgam in compliance with cur-
rent directives.
- Transmission instruments with fibre optic rod are consid-
ered resistant, while fibre optic beams may be subject to
premature wear.
- If possible, disassemble the modular instruments follow-
ing the manufacturer’s instructions and treat the individual
parts separately.
- Treat small instruments and parts in special inserts or clos-
able baskets.
- Before the automatic treatment, check that the lumen and
hollow body instruments can be treated internally and re-
move any obstructions.
- Carefully place the load in the supports.
- Objects must not be put inside each other nor cover each
other. Do not place objects so close to each other that they
cannot be washed properly.
- Arrange the load so that all surfaces can be reached by the
washing liquid. Otherwise they will not be cleaned!
- Place the objects so that the liquids can flow out without
hindrance.
- Place tall or heavy objects in the centre of the basket.
- When arranging the load, make sure that the spray arms
are not blocked by the load itself.
- Distribute the objects evenly in the baskets.
- After the treatment the transmission instruments must be
cleaned according to the manufacturer’s instructions. After
the treatment, before reusing the transmission instruments,
check that they work properly, e.g. by spraying liquid in the
sink.
When the base grid is used together with the injectors, make
sure to push the grid all the way back against the rear wall of
the tank, so that the injectors can be perfectly fixed on their
connection tube at the bottom of the tank itself.