Instructions for use - Tiva2 - Tiva2H Thermodisinfector
- Check the electrical safety according to VDE 0701/0702.
- Filter in the water flow valves.
- Replace the prefilter.
- Replace the HEPA filter.
- Check the dosing systems.
- Thermostat sensor, thermoelectric measurements.
- Conditions and functionality of the spray arms.
- Safety thermostat.
- Door gasket.
- Steam condenser.
- The functionality of the drain pump.
- Pressure switch.
- Drain pipes.
- Water flow pipes.
7.1.3 Air suction filter
The thermodisinfector is standard equipped with a suction filter
for drying air (Fig. 19), to be replaced after about 100 hours of
work (equal to about 200 cycles).
The machine can also be equipped with an «absolute» «HEPA
H14» class supplementary air filter (on request), in compliance
with EN 1822.
We recommend replacing the HEP filter every 500 working
hours, which is equivalent to 1000 operating cycles circa.