6-2
Model 3480 Electrospray Aerosol Generator
Table 6-1
Troubleshooting if the Electrospray Operation is Unstable
Problem
Solutions
Electrospray is operating in
the Dripping mode (see
Chapter 4, “Operating the
Electrospray”)
or
Current display is much
lower and more unstable
than normal
Increase the voltage setting
Increase the capillary pressure drop
Increase the air flow rate
Purge the capillary (see
Clean the capillary tip (see
Adjust the position of the capillary (see Chapter 5, “
”)
Clean the inlet fitting (see
Increase the conductivity of the sample (see
)
Electrospray is operating in
the Pulsating mode (see
Chapter 4, “Operating the
Electrospray”)
or
Current display is lower and
more unstable than normal
Increase the voltage setting
Increase the capillary pressure drop
Purge the capillary (see
)
Adjust the position of the capillary (see Chapter 5, “
”)
Clean the inlet fitting (see
Increase the conductivity of the sample (see
)
Electrospray is operating in
the Corona Discharge mode
(see Chapter 4, “Operating
the Electrospray”)
or
Current display is higher and
more unstable than normal
Decrease the voltage setting
Decrease the capillary pressure drop
Increase the CO
2
flow rate
Purge the capillary (see
Clean the capillary tip (see
Adjust the position of the capillary (see Chapter 5, “
”)
Clean the inlet fitting (see
Decrease the conductivity of the sample (see
No liquid is flowing through
the capillary
Make sure the voltage, pressure, and flow rates are adjusted properly
Make sure the capillary is immersed in the sample vial solution
”
There is no current reading
Make sure the voltage, pressure, and flow rates are adjusted properly
Make sure the capillary is immersed in the sample vial solution
Make sure the platinum wire is immersed in the sample vial solution