xiv
Model 3480 Electrospray Aerosol Generator
The Model P-2042 Nuclespot Local Air Ionizer with a Polonium-210
source must be installed in the Electrospray for it to operate properly.
Po-210 is subject to the regulations of the U.S. Nuclear Regulatory
Commission and local regulations. Carefully read the safety information
provided with the ionizer to determine your legal responsibilities regarding
the ionizer.
W A R N I N G
The use of controls, adjustments, or procedures other than those
specified in this manual may result in exposure to hazardous radiation.
Corrosive materials can degrade the protective coating of gold
on the ionizer inside the Electrospray. Do
not
operate the
Aerosol Generator with chemicals that corrode gold.
Do
not
operate the Aerosol Generator at temperatures above
50
°
C. Excess temperatures may damage the ionizer, resulting in
radioactive contamination.
The ionizer must be returned after 12 months. Contact TSI for
information and instructions on returning the ionizer.
Do
not
remove any parts from the Electrospray unless you are
specifically told to do so in this manual.
C h e m i c a l S a f e t y
Observe these warnings when operating the Electrospray Aerosol
Generator:
!
W A R N I N G
Any material or procedure mentioned in this manual is intended for use by
qualified professionals familiar with potential chemical hazards and
trained in safe laboratory procedures.
!
W A R N I N G
Corrosive materials can degrade the stainless steel and aluminum
chamber inside the Electrospray Aerosol Generator. Do
not
operate the
Electrospray with chemicals that corrode stainless steel or aluminum.