5–8
Model 3480 Electrospray Aerosol Generator
b)
Clean the optical window using a soft cloth or cotton swab
moistened with alcohol, a mild detergent, or a lens cleaner.
!
C a u t i o n
Before replacing the viewing window LED mount, make sure the optical
window is flush with the O-ring. Failure to do so could result in breakage
of the optical window as the viewing window LED mount is screwed into
the inlet manifold.
c)
Place the O-ring and optical window in the inlet manifold,
making sure the optical window is flush with the O-ring
surface, then replace the viewing window LED mount.
6.
Use a soft cloth or cotton swab moistened with alcohol or a mild
detergent to clean the inside of the aerosol exit tube and the
stainless steel surface of the aerosol exit tube that faces the
inside of the Electrospray chamber.
7.
Clean the inside of the ionization chamber using a soft cloth or
cotton swab moistened with alcohol or a mild detergent.
8.
Reassemble the inlet and exit manifolds to the ionization
chamber and connect the Electrospray chamber to the cabinet.
9.
Reattach the viewing window LED connector to the main PCB
and reconnect the three ground wires to the ionization
chamber.
10.
Place the ionizer in the ionizer retainer. The ionizer mesh-side
should be facing the open end of the ionizer retainer.
11.
Install the ionizer retainer, making sure the O-ring is between
the ionization chamber and the ionizer retainer, and tighten
with the ionizer retainer tool.
12.
Replace the inlet fitting. If needed, refer to step 11 in the section
“
” earlier in this chapter.
13.
Install the capillary tube using the instructions later in this
chapter titled “
.”
14.
Perform and air leak test using the instructions below.
Performing an Air Leak Test
Whenever an air leak is suspected or a significant part of the
Electrospray chamber has been reassembled, perform a leak test to
assure proper operation. Use these steps to check the entire
instrument for leaks. Skip to the next section if you want to isolate
the Electrospray chamber and check it for leaks.