5–14
Model 3480 Electrospray Aerosol Generator
7.
Pull the capillary straight up until it clears the high-voltage
fitting.
8.
Store the capillary in the case provided to prevent breakage.
Cleaning the Capillary Tip
If the capillary tip appears dirty, remove the capillary using the
instructions above, and rinse or soak the capillary tip in alcohol or
a mild detergent until the contaminants are removed. Then install
the capillary using the instructions in the next section.
Installing the Capillary
W A R N I N G
High voltage is accessible in several locations within this instrument.
Make sure you unplug the power source before removing the cover or
performing maintenance procedures.
To install the capillary, refer back to Figure 5-5 and use the
following instructions:
1.
Remove power from the instrument and turn off all flows.
2.
Remove the cover by loosening the four screws securing the
cover (they do not have to be removed) and pulling the cover
upward while pushing down on the top of the front panel of the
instrument or on the viewing window mount.
3.
Loosen the hex nut on the inlet fitting but do
not
remove it.
4.
Remove the high-voltage shield by loosening the mounting
screw and pulling the high-voltage shield upward.
5.
Loosen the hex nut on the high-voltage capillary guide but do
not remove it.
6.
Insert the long end of the capillary into the high-voltage
capillary guide until the end of the capillary reaches the end of
the platinum wire. Twist the capillary until it becomes retained
by the corkscrew of the platinum wire, then tighten the hex nut
on the high-voltage fitting.
7.
Attach the high-voltage shield to the cabinet by sliding the
capillary through the high-voltage shield, routing the high-
voltage wire through the groove in the bottom edge of the high-
voltage shield, and tightening the mounting screw.
8.
Turn on the power to the instrument. If the cover is removed,
high voltage will not be supplied to the instrument, but the
capillary viewing window LED will turn on.