5–10
Model 3480 Electrospray Aerosol Generator
5.
Plug all three barbed fittings.
6.
Connect a pressure gauge or manometer between the aerosol
exit port and a leak-tight valve. The valve should be closed.
7.
Connect the valve to a vacuum source or a vacuum pump
capable of drawing 18 inHg vacuum.
8.
Turn on the vacuum pump or vacuum system and slowly open
the valve. Allow the vacuum gauge reading to become stable
and then close the valve.
9.
Observe the gauge reading. The reading should not change by
more the 0.5 inHg in 1 minute.
10.
If the instrument does leak, follow the steps below under
“
Isolating Leaks
1.
Isolate the leak by wetting suspected joints with clean isopropyl
alcohol while the system is under vacuum. The alcohol will be
drawn into a leaky joint and evaporate.
2.
After isolating the leak, repair it (usually by greasing or
replacing an O-ring, or by sealing a fitting).
3.
Blow clean, dry air through the repaired section to evaporate
and remove any leftover alcohol.
4.
If you cannot find the leak, or cannot repair the leak, please
contact TSI for assistance.
Cleaning the Pressure Chamber
The pressure chamber can be cleaned if the surfaces appear dirty
or the sample vial is difficult to see clearly. Refer to Figure 5-4 and
use the following procedure to clean the pressure chamber:
W A R N I N G
High voltage is accessible in several locations within this instrument.
Make sure you unplug the power source before removing the cover or
performing maintenance procedures.
1.
Remove power from the instrument and turn off all flows.
2.
Remove the capillary using the instructions in the next section
titled “
3.
Remove the screw attaching the high-voltage cable to the fitting
on top of the pressure chamber. The fitting can then be
removed by twisting counter-clockwise, using caution not to
bend or damage the platinum high-voltage wire.