Model 3480 Specifications
A-3
Rotameter Maintenance
If you suspect either rotameter has become dirty, refer to
Figure A-2 and the instructions provided by the manufacturer
below to clean either rotameter.
!
C a u t i o n
Remove air supply pressure and bleed off output pressure prior to
performing maintenance.
RETAINING CLIP
Figure A-2
Rotameter Back View
Occasional cleaning may become necessary if
dirt appears in the flowmeter or if float
movement is restricted. To clean, rotate
“Retaining Clip” (Located in the back of
flowmeter) counter-clockwise to align clip with
slot and then pull out. Then remove top plug,
ring (if required), outlet fitting and float only.
Tube does not need to be removed. Wash tapered
hole, float, outlet fitting and top plug with clean
water and a soft brush. Rinse all parts with clean
water and dry thoroughly. Avoid use of solvents
and strong bases for cleaning. Reassemble by
reversing above.