5–12
Model 3480 Electrospray Aerosol Generator
5.
Generously grease the O-ring on the bottom half of the pressure
chamber and the pin grooves using the vacuum grease supplied
in the accessory kit. Remove excess grease and insert the
bottom half into the top half of the pressure chamber. Reapply
vacuum grease to the O-ring and pin grooves if necessary.
6.
Reinstall the high-voltage fitting, using caution not to bend or
damage the platinum high-voltage wire.
7.
Reattach the high-voltage cable to the inlet fitting and install
the capillary using the instructions in the next section titled
“
.”
S p e c i a l M a i n t e n a n c e
Removing the Capillary
The capillary needs to be removed when cleaning any part of the
Electrospray chamber or if the capillary becomes damaged or the
capillary tip becomes dirty. Use the following steps and Figure 5-5
to remove the capillary:
W A R N I N G
High voltage is accessible in several locations within this instrument.
Make sure you unplug the power source before removing the cover or
performing maintenance procedures.
1.
Remove power from the instrument and turn off all flows.
2.
Remove the cover by loosening the four screws securing the
cover (they do not have to be removed) and pulling the cover
upward while pushing down on the top of the front panel of the
instrument or on the viewing window mount.