Unpacking and Setting Up the System
2–5
!
W A R N I N G
Supplying more than 15 PSI CO
2
pressure will cause the internal silicone
tubing to disconnect from the CO
2
rotameter fitting. In addition, a higher
CO
2
inlet pressure makes it more difficult to adjust the CO
2
flow rate.
I n s t a l l i n g t h e I o n i z e r
A Model P-2042 Nuclespot Local Air Ionizer with a Polonium-210
source is shipped separately because of regulations that govern
packaging and shipping requirements for radioactive materials. The
ionizer must be installed for the Electrospray to operate properly.
Before installing the ionizer, read the safety information at the
beginning of this manual and the safety information provided with
the ionizer.
W A R N I N G
The use of controls, adjustments, or procedures other than those
specified in this manual may result in exposure to hazardous radiation.
Use this procedure to install the ionizer:
1.
Remove the ionizer retainer shown in Figure 2-2 using the
ionizer retainer tool supplied in the accessory kit, and set the
dummy source aside.
2.
Place the ionizer in the ionizer retainer in place of the dummy
source. The ionizer mesh-side should be facing the open end of
the ionizer retainer.
3.
Install the ionizer retainer, making sure the O-ring is between
the ionization chamber and the ionizer retainer, and tighten
with the ionizer retainer tool.
Note
: The ionizer must be replaced 12 months from the date on the
ionizer label. Contact TSI to order a new ionizer four weeks
before the ionizer expiration date. When returning an ionizer,
contact TSI for handling and shipping instructions.