Maintenance 5–3
Cleaning the Orifice Plate
W A R N I N G
High voltage is accessible in several locations within this instrument.
Make sure you unplug the power source before removing the cover or
performing maintenance procedures.
The orifice plate is permanently attached to the inlet fitting of the
Electrospray chamber to ensure electrical contact. Since the aerosol
is generated inside the inlet fitting and then passes through the
orifice plate, the inlet fitting will need to be cleaned more frequently
than other components through which the aerosol passes.
Figure 5-1 shows a view of the inlet fitting inside the cabinet. Use
the following steps to clean the inlet fitting and orifice plate:
Inlet Fitting
Capillary Sleeve
Screw
Electrospray Chamber
Orifice Plate
O-Ring
(2501014)
O-Ring
(2501016)
Figure 5-1
Inlet Fitting
1.
Remove power from the instrument and turn off all flows.
2.
Remove the cover by loosening the four screws securing the
cover (they do not have to be removed) and pulling the cover
upward while pushing down on the top of the front panel of the
instrument or on the viewing window mount.
3.
Remove the screw that holds the inlet fitting in place.
4.
Pull the inlet fitting out of the Electrospray chamber. Use a
twisting motion if necessary.