![TSI Incorporated 3480 Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/tsi-incorporated/3480/3480_user-manual_1178060062.webp)
5–18
Model 3480 Electrospray Aerosol Generator
High Voltage
Cable Connector
High Voltage
Module
Captive Screw (2X)
Main PCB
Figure 5-7
Reversing the High-Voltage Polarity
3.
Disconnect the high-voltage cable from the high-voltage module
by gripping the connector and twisting the collar
1
⁄
8
-turn
clockwise. The connector should then separate easily from the
socket.
4.
Loosen the two captive screws attached to the high-voltage
module until the screws pop out.
5.
Pull the high-voltage module straight out until it clears the
spacers used to secure the main PCB.
6.
Install the high-voltage module with the desired polarity in the
same orientation as the high-voltage module that was removed,
making sure the six connector pins on the main PCB are
aligned with the six sockets on the high-voltage PCB. Tighten
the two captive screws finger tight.
7.
Reconnect the high-voltage cable by pushing the connector into
the socket. There is a
1
⁄
8
-turn spring locking connect that
secures the connector when it is seated properly.
8.
Replace the cover on the instrument and tighten the four
screws that secure the cover.