Maintenance 5–15
9.
Carefully insert the capillary tip into the capillary guide. While
looking through the viewing window, continue pushing the
capillary into the capillary guide until the tip of the capillary
first appears, then tighten the hex nut on the inlet fitting.
10.
If the capillary is centered in the viewing window, installation is
complete. If not, loosen the hex nut on the inlet fitting, adjust
the position of the capillary, and retighten the hex nut.
Note:
Due to variations in the dimensions of the viewing window
components, the capillary may appear slightly off center in the
direction transverse to the capillary axis. This is not a cause
for concern however, because the center of the orifice plate is
aligned to the inlet fitting axis at the factory.
11.
Replace and secure the cabinet cover. The Electrospray is now
ready for use (See Chapter 4, “
Adjusting the Position of the Capillary Tip
If the capillary does not appear in the center of the viewing window
or you wish to move the capillary closer or further from the orifice
plate, refer to the instruction preceding this subsection titled,
“
.” Step 10 lists instructions for adjusting
the position of the capillary tip. Make sure to remove power from
the instrument and turn off all flows before removing the cover from
the instrument.
Adjusting the Position of the Capillary and the
Platinum Wire in the Pressure Chamber
In the pressure chamber, the capillary and the platinum wire are
positioned at the factory to extend near the bottom of a standard
sample vial (1.7mL centrifuge tube). In most cases, the sample to
be electrosprayed can be put into a standard sample vial; however,
if a smaller vial must be used, the following things must be done:
C a u t i o n
The electronic circuits within this instrument are susceptible to
electrostatic discharge (ESD) damage. Use ESD precautions to avoid
damage.
Use only a table top with a grounded conducting surface.
Wear a grounded, static-discharging wrist strap.
W A R N I N G
High voltage is accessible in several locations within this instrument.
Make sure you unplug the power source before removing the cover or
performing maintenance procedures.