Maintenance 5–7
Aerosol
Exit Tube
Exit Manifold Block
O-Ring (2501029)
Ionization Chamber
Viewing Window
Lens Assembly
Inlet Manifold Block
Hex Nut
Capillary Inserts
Here
Inlet Fitting
Hex Nut
Common Ground Screw
Figure 5-3
Cleaning the Electrospray Chamber
3.
Remove the common ground screw on the side of the
Electrospray chamber. Also detach the viewing window LED
connector from the main PCB.
4.
Detach the Electrospray chamber from the cabinet by removing
the two screws on the back panel of the instrument behind the
Electrospray chamber (not shown in Figure 5-3).
5.
Remove the eight screws that fasten the two white manifold
blocks to the inlet and exit of the ionization chamber. The
viewing window lens can be cleaned if necessary by following
the instructions in the section “
” earlier in this chapter. The viewing window LED mount
can also be removed to clean the optical window it holds in
place by using the following procedure:
a)
Unscrew the viewing window LED mount. The optical
window and an O-ring will fall out of the inlet manifold.