5–4
Model 3480 Electrospray Aerosol Generator
5.
Detach the capillary from the inlet fitting (see “
6.
Remove the O-rings using a tweezers or similar tool. Inspect the
O-rings for damage and replace if necessary.
7.
Use a soft cloth soaked in alcohol or a mild solvent to clean the
outside of the inlet fitting and flush the inside of the inlet fitting
with alcohol or a mild solvent to clear any deposits that may
have formed.
8.
Rinse the inlet fitting with clean water.
9.
Blow clean, dry air through and on the inlet fitting to evaporate
the water.
10.
Replace the O-rings and lightly apply the vacuum grease
supplied with the instrument. Remove any excess grease.
11.
Replace the inlet fitting and reinstall the capillary (see
“
” later in this chapter).
12.
Replace the cover on the instrument and tighten the four
screws that secure the cover.
Cleaning the Aerosol Exit Port
The aerosol exit port can be cleaned as needed without
disassembling the instrument. Simply remove power and turn off
all flows to the instrument and use a cotton swab moistened with
alcohol or a mild detergent to clean the inside of the aerosol exit
port (see Figure 3-2).
Cleaning the Viewing Window Lens
If the viewing window lens becomes dirty, it can be cleaned without
removing the cover from the instrument. First try cleaning the lens
without disassembling the instrument by using a cotton swab
soaked in alcohol, a mild detergent, or a lens cleaner. If the lens is
still dirty, use the following procedure to clean the viewing window
lens:
1.
Remove power from the instrument and turn off all flows.
2.
Using Figure 5-2 as a guide, unscrew the viewing window
mount by hand and remove the lens ring retainer using a coin.
3.
Turn the viewing window mount upside down to remove the
lens.
4.
Clean the lens using a soft cloth soaked in alcohol, a mild
detergent, or a lens cleaner.
5.
If necessary, the O-rings can be removed from the viewing
window mount and the viewing window mount can be cleaned
using a soft cloth soaked in alcohol or a mild detergent.