A-2
Model 3480 Electrospray Aerosol Generator
Environmental Conditions
...
Indoor use
Altitude up to 2000 m (6500 ft)
Ambient temperature 10-50
°
C
Ambient humidity 0-90% RH non-condensing
Over-voltage category II
Pollution degree II
*TSI Incorporated is authorized by the United States Nuclear Regulatory Commission to distribute
these Ionizers. If your location is within the United States, no other federal license is required. Check
local regulations for your own protection. Ionizers are shipped separately. End-user name and
address is required. Contact TSI to return used ionizers or obtain replacements.
Specifications are subject to change without notice. TSI and the TSI logo are registered trademarks of
TSI Incorporated.
Pressure Regulator Maintenance
The air supplied to the pressure regulator should always be from a
clean, dry air supply, eliminating the need to clean the pressure
regulator. However, if you suspect the regulator has become dirty,
use Figure A-1 and the instructions provided by the manufacturer
below to disassemble and clean the regulator.
!
C a u t i o n
Remove air supply pressure and bleed off output pressure prior to
performing maintenance.
Figure A-1
Pressure Regulator Assembly
To disassemble the pressure regulator, loosen
adjusting handwheel (1) until spring tension is
relieved. Remove six screws (9) and lift off bonnet
(5), spring plate (6), spring (7) and diaphragm
assembly (8). To remove nozzle assembly (10),
use
5
⁄
8
” socket wrench to remove nozzle assembly
to avoid damage to the nozzle. All parts may be
cleaned by immersion in a suitable solvent and
blowing dry with air stream.
To reassemble, replace nozzle assembly (10) and
place diaphragm assembly (8) over body (11),
with diaphragm plate up. Place spring (7) and
spring plate (6) on diaphragm assembly (8), re-
install bonnet (5) and tighten six screws (9). The
six screws should be tightened alternately.