Controls, Indicators, and Connectors
3–5
Analog Output
The analog output is a standard 15-pin D-sub connection that
allows analog signals to be read by an external instrument to
collect Electrospray voltage and current information. The pin
designations and signal connections are shown in Figure 3-3 and
Table 3-1, respectively. Note the conversion factors in Table 3-1.
5 4 3 2 1
9
8 7 6
10
11
12
13
14
15
Figure 3-3
ANALOG OUTPUT Pin Designations
Table 3-1
Signal Connections for Analog Output Configurations
Pin Number(s)
Signal
1
Voltage Output (1V = 1.00 kV)
2
Voltage
GND
3
Current Output (1V = 100 nA)
4
Current
GND
5-15
—
AC Power Connector
The AC power connector accepts the line cord (supplied) to provide
AC power to the instrument. Line voltage can be 85 to 260 VAC,
50/60 Hz, single phase, 3A max. The connector has a built-in
on/off switch. Power consumption is 25 Watts, max.
Note
: Make certain the line cord is plugged into a grounded power
outlet. Position the Electrospray so the power cord connector is
easily accessible.
Filtered Air Inlet
The filtered air inlet connection is described in Chapter 2,
“
Unpacking and Setting Up the System
.”
CO
2
Inlet
The CO
2
inlet connection is described in Chapter 2, “
.”