EN
5
5.
Attach one end of the supplied
EZ Twist
TUBING
to the fitting on the bottom of the nebulizer and the other end
to your compressed air source. Ensure both ends are securely engaged.
NEBULIZER TREATMENT
The reusable
AEROECLIPSE
* XL BAN can be operated in
CONTINUOUS MODE
(constantly producing aerosol) by
rotating the
MODE SELECTOR
to the
CONTINUOUS MODE
position (
– Fig. 2). The nebulizer can be returned to
BREATH
ACTUATED
Mode by rotating the
MODE SELECTOR
to the
BREATH ACTUATED
position (
– Fig. 2).
UÊ Ü>ÞÃÊ vÜÊ ÞÕÀÊ i>Ì V>ÀiÊ «ÀviÃÃ>½ÃÊ ÌÀi>ÌiÌÊ
instructions.
UÊ / iÊÀiÕÃ>LiÊ
AEROECLIPSE
* XL BAN should be used with
the
Ombra
* Table Top or Portable Compressors or an air
source capable of delivering a flow of 3.5 to 5 litres per minute
(Lpm) under an operating pressure of 0.92 to 1.79 bar (13.3 to
26 pounds per square inch [psi]).
UÊ 7 iÊ Ì iÊ `iÛViÊ ÃÊ Li}Ê ÕÃi`Ê ÌÊ `iÛiÀÊ >iÀÃÊ ÌÊ ÞÕ}Ê
children or anyone requiring assistance, they should be
supervised at all times to ensure a safe and effective treatment.
BREATH ACTUATED MODE
Ensure the
MODE SELECTOR
is in
BREATH ACTUATED
position (
– Fig. 2).
1.
While sitting in a comfortable, upright position, switch on your compressor.
2.
Place the
MOUTHPIECE
in your mouth, ensuring your lips are sealed around the edge. Be careful not to
cover the
EXHALATION VALVE
on the bottom of the
MOUTHPIECE
, as doing so will prevent the valve from
functioning properly.
3.
Inhale slowly and deeply. As you inhale, the
GREEN FEEDBACK BUTTON
on the top of the nebulizer (Fig. 1)
will move down, indicating that the reusable
AEROECLIPSE
* XL BAN is producing aerosol in response to
your inhalation.
4.
Exhale normally through the device. The valve on the mouthpiece only opens during exhalation to allow air
to exit the device. This helps prevent re-breathing of exhaled air. During exhalation, the
GREEN FEEDBACK
BUTTON
returns to the up position indicating no aerosol is being produced (Fig. 7).
5.
Continue to breathe normally through the device until it begins to make a ‘sputter’ noise or as instructed by your
healthcare professional.
6.
Switch off the compressor when finished with your treatment.
NOTE
: If you are unable to trigger the device during inhalation (unable to get the
GREEN FEEDBACK
BUTTON
to go down when you breathe in) it can be manually operated with your index finger by depressing/
releasing the
GREEN FEEDBACK BUTTON
in coordination with your breathing. Alternately, you may switch
the device to
CONTINUOUS MODE
(
– Fig. 2).
CONTINUOUS MODE
Ensure the
MODE SELECTOR
is in
CONTINUOUS
position (
– Fig. 2).
In
CONTINUOUS
Mode, the nebulizer may be used with the reusable
ComfortSeal
* Mask. Refer to ‘Setting Up
Your Nebulizer’, Step 3 on how to properly use a mask with the nebulizer.
1.
While sitting in a comfortable, upright position, switch on your compressor.
Continuous Mode ‘down’ position
Breath Actuated Mode ‘up’ position
FIGURE 7