![Trudell Medical International AeroEclipse XL Скачать руководство пользователя страница 6](http://html.mh-extra.com/html/trudell-medical-international/aeroeclipse-xl/aeroeclipse-xl_owners-manual_1174020006.webp)
EN
4
position (
– Fig. 2). If the
GREEN FEEDBACK BUTTON
still does not move freely contact your authorized
distributor for a new unit.
2.
Examine the
MOUTHPIECE
to ensure the
EXHALATION VALVE
is on the bottom, facing down to ensure
exhaled breath is directed away from the face.
3.
For patients unable to use the
MOUTHPIECE,
the reusable
AEROECLIPSE
* XL BAN can be used with an
optional Mask by removing the
MOUTHPIECE
and inserting the Mask into the
MOUTHPIECE
opening
(Fig. 3). The reusable
ComfortSeal
* Mask (Fig. 4) has an
ELBOW ADAPTER
that can be inserted directly
into the Mouthpiece opening.
NOTE:
When using the reusable
AEROECLIPSE
* XL BAN with a
ComfortSeal
* Mask, ensure a secure
seal all around the edge of the Mask. Even tiny gaps may allow ambient air to enter the Mask, significantly
reducing the amount of medication delivered.
FIGURE 6
FIGURE 5
FIGURE 3
FIGURE 4
Exhalation Valve
Elbow Adapter
4.
Unscrew and remove the
NEBULIZER TOP
. Carefully place prescribed medication into
NEBULIZER CUP
(Fig. 5). The reusable
AEROECLIPSE
* XL BAN has a minimum fill of 1 mL and a maximum fill of 6 mL.
Reattach the
NEBULIZER TOP
, making sure to line up the
WHITE CYLINDER
with the
STEM
of the
NEBULIZER CUP
and gently hand tighten until a stop is detected, as shown in Figure 6. Ensure nebulizer is
kept in an upright position so medication does not spill. (e.g. Use your compressor’s nebulizer cup holder).