Maintenance and Troubleshooting
Thermo Fisher Scientific
Maintenance and Troubleshooting 4-17
REC 4132 Rev
E
If the alignment of the touch screen is off making it difficult to use the
screen it can be recalibrated.
Path:
More Menus/Utility/Touch Screen Calibration
The following screen displays:
Figure 4–11. Touch Screen Calibration Screen
The screen asks the user to press three targets on the screen, indicated by
a large plus sign ‘+’. The first target is in the upper left of the screen, the
second is in the lower right of the screen and the third is in the upper
right of the screen.
If the touch screen is so out of alignment the user is unable to login, the
following procedure starts the touch screen calibration utility.
1.
Power up the machine
2.
When the User Login screen displays, start the belt
3.
What for the code
TSC-1
to appear next to
MORE MENUS
4.
Stop the belt
5.
Wait for the code
TSC-2
to display
6.
Start the belt again
7.
Wait for the code
TSC-3
to display
8.
Stop the belt again, this should trigger the touch screen calibration
Note.
There is a 1 minute time limit when the EZx starts where
this procedure can be used to start a screen calibration.
▲
Touch Screen
Calibration
Содержание EZx
Страница 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Страница 7: ......
Страница 8: ......
Страница 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Страница 14: ......
Страница 20: ...Contents vi Contents Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 40: ...Introduction 1 8 Introduction Thermo Fisher Scientific REC 4132 Rev E A96255 Figure 1 5 Drop Down Selections...
Страница 80: ...Operation 3 32 Operation Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 112: ......
Страница 203: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 3 REC 4132 Rev E...
Страница 204: ...Critical Radiation Survey C 4 Critical Radiation Survey Thermo Fisher Scientific REC 4132 Rev E...
Страница 205: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 5 REC 4132 Rev E...
Страница 213: ...Filter Selection Thermo Fisher Scientific Filter Selection D 5 REC 4132 Rev E This Page Left Blank Intentionally...
Страница 214: ......
Страница 218: ...Package Edge Padding E 4 Package Edge Padding Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 222: ...Network Settings F F 4 Network Settings Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 226: ...Statistics File G 2 Statistics File Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 228: ......
Страница 231: ...This Page Left Blank Intentionally...
Страница 232: ......
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......