Installation
2-6
Installation
Thermo Fisher Scientific
REC 4132 Rev
E
Line power should be supplied through one of the holes provided on the
right side of the lower enclosure. The proper wiring should be capable of
providing 15A/100 VAC or 7.5A/240 VAC.
EZx with a Vortex cooler 14 AWG stranded wire should be used with
watertight metal conduit. EZx with an air conditioner 12 AWG stranded
wire should be used with watertight metal conduit. The proper wiring
should be capable of providing 20A/115VAC or 10A/230 VAC. The
branch circuit must be externally fused at no more than 20A.
The line power connection terminals for the
EZx
are located on the
wiring panel terminal block TB1. The “hot” and “neutral” conductors
must be connected to terminals L1 and L2 respectively. These terminals
have a torque specification of 0.6 N/m or 5.5 lbs/in.
A ground terminal located on the bottom left of the enclosure is
provided for safety ground connections. This terminal should be
tightened to 5.1 N/m or 45 lb/in.
Refer to the “
Back Panel Wiring Diagram
” (
Appendix F
) for detailed
information. All Line power wiring must conform to National Electric
Code (US) and/or any local regulations. The EZx incorporates universal
voltage and connects anywhere in the world.
A96733
Figure 2–3.
I/O Board
Connecting
Power to the
System
A96732
Ground
Содержание EZx
Страница 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Страница 7: ......
Страница 8: ......
Страница 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Страница 14: ......
Страница 20: ...Contents vi Contents Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 40: ...Introduction 1 8 Introduction Thermo Fisher Scientific REC 4132 Rev E A96255 Figure 1 5 Drop Down Selections...
Страница 80: ...Operation 3 32 Operation Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 112: ......
Страница 203: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 3 REC 4132 Rev E...
Страница 204: ...Critical Radiation Survey C 4 Critical Radiation Survey Thermo Fisher Scientific REC 4132 Rev E...
Страница 205: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 5 REC 4132 Rev E...
Страница 213: ...Filter Selection Thermo Fisher Scientific Filter Selection D 5 REC 4132 Rev E This Page Left Blank Intentionally...
Страница 214: ......
Страница 218: ...Package Edge Padding E 4 Package Edge Padding Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 222: ...Network Settings F F 4 Network Settings Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 226: ...Statistics File G 2 Statistics File Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 228: ......
Страница 231: ...This Page Left Blank Intentionally...
Страница 232: ......
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......