Thermo Fisher Scientific
REC 4261 Rev C PN 091811
English
August 2007
EZx
™
Contaminant Detector
Quick Start Procedures
Congratulations on the purchase of your
EZx
Contaminant Detector
. This
Quick Start Guide
is
intended to get your detector up and running. For full
operational procedures related to the
EZx
, refer to your
User Guide
(REC 4132).
The following steps have to be performed or configured
for the best performance, reliability, and safety.
•
Long Seasoning of the Tank
– preconditions the X-
ray tube and the tank environment.
•
Radiation Survey
– always perform a Radiation
Survey if this is the first time the
EZx
has been
operated after installation.
•
New Product Setup
– includes
Reject
,
Filter
, and
Shift
setup.
S
TART
UP
At start up the following screen displays:
1.
Select
Username,
System Administrator
2.
Enter
Thermo1
as the password
The Main Screen displays:
L
ONG
S
EASONING
Select
More Menus/Test/Tube Diagnostics/Start Long
Season.
The EZx indicates when
Long
seasoning is
complete.
R
ADIATION
S
URVEY
During the
Long Seasoning
, when the current X-ray
power supply KV and mA are above ~65 kV and ~w.5
mA perform a Radiation Survey. (The current readings
are displayed in the center of the screen.).
Using an approved radiation detector check all areas of
the machine indicated on the
Radiation Survey
Sheet
(
Appendix C in REC 4132
).
R
EJECT
S
ETUP
1.
Select
More Menus/Setup/IO/Output
2.
Enter
Distance
– how far the reject device is from
the X-rays, in the center of the aperture tunnel)
3.
Enter
Duration
– how long the signal to the reject
device should be on.
4.
Select
Previous
5.
Select
Close
N
EW
P
RODUCT AND
F
ILTER
S
ETUP
1.
Select
More Menus/Product Setup/New Product
2.
Enter
Package Length
3.
Enter
Belt Speed
4.
Enter the minimum
Package Gap
(distance between
packages).
5.
Select
Quick Learn
Содержание EZx
Страница 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Страница 7: ......
Страница 8: ......
Страница 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Страница 14: ......
Страница 20: ...Contents vi Contents Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 40: ...Introduction 1 8 Introduction Thermo Fisher Scientific REC 4132 Rev E A96255 Figure 1 5 Drop Down Selections...
Страница 80: ...Operation 3 32 Operation Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 112: ......
Страница 203: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 3 REC 4132 Rev E...
Страница 204: ...Critical Radiation Survey C 4 Critical Radiation Survey Thermo Fisher Scientific REC 4132 Rev E...
Страница 205: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 5 REC 4132 Rev E...
Страница 213: ...Filter Selection Thermo Fisher Scientific Filter Selection D 5 REC 4132 Rev E This Page Left Blank Intentionally...
Страница 214: ......
Страница 218: ...Package Edge Padding E 4 Package Edge Padding Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 222: ...Network Settings F F 4 Network Settings Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 226: ...Statistics File G 2 Statistics File Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 228: ......
Страница 231: ...This Page Left Blank Intentionally...
Страница 232: ......
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......