EZx Menus
Thermo Fisher Scientific
EZx Menus
A-43
REC 4132
Rev E
Hint.
Below are a few helpful hints for adjusting filter
thresholds.
▲
•
If the product changes density during the day, these
thresholds may need adjustment.
•
Increasing the filter threshold values decreases the sensitivity.
Decreasing the values increases the sensitivity that in turn
increases the likelihood of false rejects.
Path: More Menus/Product Setup/Edit Product/Power Supply
When a
Quick Learn
is performed, the
EZx
sets default power settings
based on the aperture size and the belt speed. The higher the mA, the
brighter the image, the KV increases the
X-ray
penetration. If after a
Quick Learn
the product image is not clear, change the settings here.
Note.
If the Power Supply settings and/or the Filter Settings are
setup manually, running a Quick Learn resets them to the default
and they need to be manually set again.
▲
Hint.
If the product image is completely washed out and the KV is
decreased the product displays darker. The goal is to be able to
see some of the product.
▲
Power Supply Setup
Содержание EZx
Страница 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Страница 7: ......
Страница 8: ......
Страница 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Страница 14: ......
Страница 20: ...Contents vi Contents Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 40: ...Introduction 1 8 Introduction Thermo Fisher Scientific REC 4132 Rev E A96255 Figure 1 5 Drop Down Selections...
Страница 80: ...Operation 3 32 Operation Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 112: ......
Страница 203: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 3 REC 4132 Rev E...
Страница 204: ...Critical Radiation Survey C 4 Critical Radiation Survey Thermo Fisher Scientific REC 4132 Rev E...
Страница 205: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 5 REC 4132 Rev E...
Страница 213: ...Filter Selection Thermo Fisher Scientific Filter Selection D 5 REC 4132 Rev E This Page Left Blank Intentionally...
Страница 214: ......
Страница 218: ...Package Edge Padding E 4 Package Edge Padding Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 222: ...Network Settings F F 4 Network Settings Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 226: ...Statistics File G 2 Statistics File Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 228: ......
Страница 231: ...This Page Left Blank Intentionally...
Страница 232: ......
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......