Operation
3-16 Operation
Thermo Fisher Scientific
REC 4132 Rev
E
2.
Enter a
Name
for the new product. (Refer to
Data Entry
on page 1-
6)
3.
In the same manner, enter the
Package Length
4.
Enter the
Belt Speed
(the default belt speed is based on the aperture
size and is the ideal speed for your setup)
5.
Enter the minimum
Package Gap
(the minimum distance between
packages)
6.
Package Detection
is used for packages the photoeye might not see
completely. As an example a sealed package that might have a
leading and trailing flap. Put the measurements of the package flap
in the
Pad leading edge
and
Pad trailing edge
in the
Package
Detection
field. See Appendix E for additional information on
Pad
Leading Edge
and
Pad Trailing Edge
.
Note.
If a product is not completely setup the product icon next to
the product name has a red X and that product cannot be run until
setup is complete.
▲
Path
: More Menus/Product Setup/(New) Edit Product/Reject Setup
The
P
RODUCT
R
EJECTOR
setup determines the
Distance
(how far the
reject device is from the X-ray),
Duration
(how long the signal to the
reject device should be on), and at what point on the package the rejecter
should fire (either the front or center of the package).The
Reject
Package Signal Output
is setup in the same manner as the reject. It can
be used to activate an external reject indicator. At this time, it is used to
tell a checkweigher not to weigh a package and not to reject it.
Distance
(distance to the checkweigher) and
Duration
(how long the signal to the
checkweigher should be on).
Reject Setup
Содержание EZx
Страница 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Страница 7: ......
Страница 8: ......
Страница 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Страница 14: ......
Страница 20: ...Contents vi Contents Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 40: ...Introduction 1 8 Introduction Thermo Fisher Scientific REC 4132 Rev E A96255 Figure 1 5 Drop Down Selections...
Страница 80: ...Operation 3 32 Operation Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 112: ......
Страница 203: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 3 REC 4132 Rev E...
Страница 204: ...Critical Radiation Survey C 4 Critical Radiation Survey Thermo Fisher Scientific REC 4132 Rev E...
Страница 205: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 5 REC 4132 Rev E...
Страница 213: ...Filter Selection Thermo Fisher Scientific Filter Selection D 5 REC 4132 Rev E This Page Left Blank Intentionally...
Страница 214: ......
Страница 218: ...Package Edge Padding E 4 Package Edge Padding Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 222: ...Network Settings F F 4 Network Settings Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 226: ...Statistics File G 2 Statistics File Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 228: ......
Страница 231: ...This Page Left Blank Intentionally...
Страница 232: ......
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......