![ThermoFisher Scientific EZx Скачать руководство пользователя страница 73](http://html1.mh-extra.com/html/thermofisher-scientific/ezx/ezx_user-manual_1105591073.webp)
Operation
Thermo Fisher Scientific
Operation 3-25
REC 4132 Rev
E
A96644
Figure 3–24.
Other Devices Screen
At this screen, the setup is the normal state of the input
indicating everything is operating properly. If the device changes
from the normal state then something is not working properly
and an alarm is activated.
EXAMPLE:
If air pressure is setup and the air pressure changes
state (something goes wrong i.e. pressure drops) the air pressure
alarm is activated. The external alarm label is reported in the log
when the alarm is active.
The
EZx
has the capacity to configure 7 outputs with 10 output
assignments. Output 8, the Cooling Device, cannot be un-assigned or be
configured to another output type.
Note:
The default outputs for the EZx software are Output 1
assigned to Reject Verify, Output 2 assigned Alarm, and Output 8
assigned Cooling Device.
▲
Output Setup
Содержание EZx
Страница 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Страница 7: ......
Страница 8: ......
Страница 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Страница 14: ......
Страница 20: ...Contents vi Contents Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 40: ...Introduction 1 8 Introduction Thermo Fisher Scientific REC 4132 Rev E A96255 Figure 1 5 Drop Down Selections...
Страница 80: ...Operation 3 32 Operation Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 112: ......
Страница 203: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 3 REC 4132 Rev E...
Страница 204: ...Critical Radiation Survey C 4 Critical Radiation Survey Thermo Fisher Scientific REC 4132 Rev E...
Страница 205: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 5 REC 4132 Rev E...
Страница 213: ...Filter Selection Thermo Fisher Scientific Filter Selection D 5 REC 4132 Rev E This Page Left Blank Intentionally...
Страница 214: ......
Страница 218: ...Package Edge Padding E 4 Package Edge Padding Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 222: ...Network Settings F F 4 Network Settings Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 226: ...Statistics File G 2 Statistics File Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 228: ......
Страница 231: ...This Page Left Blank Intentionally...
Страница 232: ......
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......