About this Manual
viii
About this Manual
Thermo Fisher Scientific
REC 4132 Rev
E
Before the X-ray machine can be installed and operated on the work
premises in Canada, the following actions must be taken:
•
Contact your federal/provincial/territorial radiation protection
authority and give advanced notice of the installation.
•
Within 48 hours of installation, a radiation leakage test must be
performed by a qualified person to ensure the regulatory limit is not
exceeded. (0.5 mR per hour at 5.0 cms from all external surfaces of
the device.)
•
Prepare, sign, and date a radiation report (survey) that states
quantitatively the radiation levels measured on all surfaces of the
machine.
•
Attach a copy of the validated calibration certificate for the survey
meter used and forward both to your federal/provincial/territorial
authority.
•
Retain a copy of both documents for your records.
Before any
X-ray
machine can be installed and operated on the work
premises in the United States, the following actions must take place:
•
Contact your State Health Department Authority and give advance
notice of the planned installation;
•
Follow the rules outlined by state, government, and local authorities
for radiation examination; and
•
A
Critical Radiation Survey
must be carried out and recorded by a
qualified person to certify the machine is safe to use (usually a
Thermo Fisher Scientific
Service Technician.)
X-ray
s are part of the electromagnetic radiation spectrum. This spectrum
includes radio waves, microwaves, infrared, the visible spectrum, ultra
violet and gamma radiation.
X-ray
s are a short wavelength of the order
of 10 - 10
10
meters.
X-ray
s are also called ionizing radiation because of
the way they interact with matter.
X-ray
s are produced in the
EZx
by means of a cathode ray tube, which is
similar to the tube of a television. Electrons hit a metal target at high
speed and generate
X-ray
s. Over 95% of this energy is converted to heat.
Canadian
Requirements for
Installation/Operation
US Requirements for
Installation/Operation
Production
and Nature of
X-ray
s
Содержание EZx
Страница 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Страница 7: ......
Страница 8: ......
Страница 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Страница 14: ......
Страница 20: ...Contents vi Contents Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 40: ...Introduction 1 8 Introduction Thermo Fisher Scientific REC 4132 Rev E A96255 Figure 1 5 Drop Down Selections...
Страница 80: ...Operation 3 32 Operation Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 112: ......
Страница 203: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 3 REC 4132 Rev E...
Страница 204: ...Critical Radiation Survey C 4 Critical Radiation Survey Thermo Fisher Scientific REC 4132 Rev E...
Страница 205: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 5 REC 4132 Rev E...
Страница 213: ...Filter Selection Thermo Fisher Scientific Filter Selection D 5 REC 4132 Rev E This Page Left Blank Intentionally...
Страница 214: ......
Страница 218: ...Package Edge Padding E 4 Package Edge Padding Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 222: ...Network Settings F F 4 Network Settings Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 226: ...Statistics File G 2 Statistics File Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 228: ......
Страница 231: ...This Page Left Blank Intentionally...
Страница 232: ......
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......