Operation
Thermo Fisher Scientific
Operation 3-9
REC 4132 Rev
E
Check the emergency stop to ensure it is working correctly. The X-rays
and the conveyor belt motor are disengaged when an emergency stop is
activated. When the E-stop switch is re-set, X-rays should not come
back on and the motor should not start until the machine is instructed to
do so using the touch screen and the motor start/stop switch.
Path
: More Menus/Test/X-ray Diagnostics
1.
As in the Critical Radiation Survey select
START
X-
RAYS
2.
After the X-rays have started, push the
EMERGENCY STOP
located
on the front of the machine and make sure the X-rays disengage.
Check that the interlocks and E-stops are working correctly. The
interlocks are placed inside the conveyor hoods and prevent X-rays from
being emitted once the hood is slightly open. The E-stop is to be used in
case of emergency to stop the x-rays and conveyor belt.
Check the
Emergency
Stop
Check
Interlocks and
E-Stop
Bottom Panel
Front Panel
Center Hood
Left and Right Hoods
E-Stop
Содержание EZx
Страница 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Страница 7: ......
Страница 8: ......
Страница 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Страница 14: ......
Страница 20: ...Contents vi Contents Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 40: ...Introduction 1 8 Introduction Thermo Fisher Scientific REC 4132 Rev E A96255 Figure 1 5 Drop Down Selections...
Страница 80: ...Operation 3 32 Operation Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 112: ......
Страница 203: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 3 REC 4132 Rev E...
Страница 204: ...Critical Radiation Survey C 4 Critical Radiation Survey Thermo Fisher Scientific REC 4132 Rev E...
Страница 205: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 5 REC 4132 Rev E...
Страница 213: ...Filter Selection Thermo Fisher Scientific Filter Selection D 5 REC 4132 Rev E This Page Left Blank Intentionally...
Страница 214: ......
Страница 218: ...Package Edge Padding E 4 Package Edge Padding Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 222: ...Network Settings F F 4 Network Settings Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 226: ...Statistics File G 2 Statistics File Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 228: ......
Страница 231: ...This Page Left Blank Intentionally...
Страница 232: ......
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......