![ThermoFisher Scientific EZx Скачать руководство пользователя страница 82](http://html1.mh-extra.com/html/thermofisher-scientific/ezx/ezx_user-manual_1105591082.webp)
Maintenance and Troubleshooting
4-2 Maintenance and Troubleshooting
Thermo Fisher Scientific
REC 4132 Rev
E
•
Covers over the electronics and moving parts must always
remain in place during normal operation. Remove only for
maintenance with the machine’s power OFF. Replace all
covers before resuming operation.
•
During maintenance, a safety tag (not supplied by
Thermo
Fisher Scientific
) is to be displayed over the ON/OFF switch
area instructing others not to operate the unit (ANSI:B157.1)
If the system has an air conditioner that replaced the Vortex cooler,
follow the AC manufactures’ recommendation for cleaning the air filter
on the AC. Refer to the instruction manual that came with the AC.
The
EZx
has been designed to be as maintenance-free as possible. Wear
items such as the conveyor belt or rollers need to be replaced regularly.
Interlocks and E-stops need to be replaced if they are damaged or
broken.
Preventative daily maintenance by of the operator ensures the
EZx
continues to run normally. Some of the weekly and monthly tasks
require maintenance staff and training. Quarterly and yearly service
should be provided by a
Thermo Fisher Scientific
representative.
The following daily tasks are recommended prior to starting up the
production line each day.
•
Clean the system if necessary
•
Perform a routine
Critical Radiation Survey
of the system as
required by government, state, and local authorities. Record the
results in a log file for the system (see Appendix A).
Note.
Radiation Surveys should be performed in accordance with
your government, state, and local authorities, which may not be
daily.
▲
•
Check the condition of the shielding curtains at the entry and exit of
the hood.
•
Check the belt tension
•
Check the air pressure
Systems with
Air
Conditioners
Preventative
Maintenance
Daily
Maintenance
Содержание EZx
Страница 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Страница 7: ......
Страница 8: ......
Страница 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Страница 14: ......
Страница 20: ...Contents vi Contents Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 40: ...Introduction 1 8 Introduction Thermo Fisher Scientific REC 4132 Rev E A96255 Figure 1 5 Drop Down Selections...
Страница 80: ...Operation 3 32 Operation Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 112: ......
Страница 203: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 3 REC 4132 Rev E...
Страница 204: ...Critical Radiation Survey C 4 Critical Radiation Survey Thermo Fisher Scientific REC 4132 Rev E...
Страница 205: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 5 REC 4132 Rev E...
Страница 213: ...Filter Selection Thermo Fisher Scientific Filter Selection D 5 REC 4132 Rev E This Page Left Blank Intentionally...
Страница 214: ......
Страница 218: ...Package Edge Padding E 4 Package Edge Padding Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 222: ...Network Settings F F 4 Network Settings Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 226: ...Statistics File G 2 Statistics File Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 228: ......
Страница 231: ...This Page Left Blank Intentionally...
Страница 232: ......
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......