Maintenance and Troubleshooting
Thermo Fisher Scientific
Maintenance and Troubleshooting 4-13
REC 4132 Rev
E
5.
When the
Diode Calibration
is complete, the machine indicates if
the calibration has passed or failed.
If the
Diode Calibration
fails, run the calibration again. If it fails for
a second time, contact
Thermo Fisher Scientific Field Service
.
6.
When the calibration is complete and has passed, begin running
product as usual.
Table 4–2.
EZx
Weekly Maintenance
Weekly Maintenance
Completed
Check interlocks and E-stops
See page 4-8
Check the oil temperature in the tank
See page 4-8
Diode Calibration
See page 4-11
CAUTION.
Do not use the interlocks or E-stops as the normal
way to stop
X-ray
s. The usable life of the
X-ray
generator will be
shortened if this is done on a regular basis.
On a monthly basis, do the following:
•
Check roller bearings for excessive noise or vibration. Replace as
necessary.
•
Check to see the annunciation lamp is working correctly (
Figure 4–
8
)
•
Check inside the machine for presence of any oil leaks from the
tank. If there is oil from the tank contact
Thermo Fisher Scientific
immediately.
Monthly
Maintenance
Содержание EZx
Страница 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Страница 7: ......
Страница 8: ......
Страница 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Страница 14: ......
Страница 20: ...Contents vi Contents Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 40: ...Introduction 1 8 Introduction Thermo Fisher Scientific REC 4132 Rev E A96255 Figure 1 5 Drop Down Selections...
Страница 80: ...Operation 3 32 Operation Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 112: ......
Страница 203: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 3 REC 4132 Rev E...
Страница 204: ...Critical Radiation Survey C 4 Critical Radiation Survey Thermo Fisher Scientific REC 4132 Rev E...
Страница 205: ...Critical Radiation Survey Thermo Fisher Scientific Critical Radiation Survey C 5 REC 4132 Rev E...
Страница 213: ...Filter Selection Thermo Fisher Scientific Filter Selection D 5 REC 4132 Rev E This Page Left Blank Intentionally...
Страница 214: ......
Страница 218: ...Package Edge Padding E 4 Package Edge Padding Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 222: ...Network Settings F F 4 Network Settings Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 226: ...Statistics File G 2 Statistics File Thermo Fisher Scientific REC 4132 Rev E This Page Left Blank Intentionally...
Страница 228: ......
Страница 231: ...This Page Left Blank Intentionally...
Страница 232: ......
Страница 235: ......
Страница 236: ......
Страница 237: ......
Страница 238: ......
Страница 239: ......
Страница 240: ......