3
OMNICxi Software Operation
Verify Instrument Performance
Thermo Scientific
DXRxi User Guide
167
If all the test results show “Passed” as in the example above, the instrument passed the PV
test. If one or more tests show a “fail” result, refer to the
information
for help.
10. When you are finished viewing the results, click
Close
to close the window.
The software automatically saves the measured PV data (polystyrene spectrum and image
map) in this location:
Thermo/OMNICxi/Spectra/PV Data
To view PV test results
1. Click the PV status icon and choose
Results
.
A window opens with results from the most recent PV test for the installed laser set. Here
is an example: