System Tools
158
DXRxi User Guide
Thermo Scientific
8. Locate the pinhole in the substrate.
Turn the brightness control most of the way down (the pinhole is backlit). Adjust the X
and Y position of the stage until the substrate fills the field of view. Then look for the
pinhole. (It is not necessarily in the center of the opening in the alignment tool.)
Use the focus control to bring the pinhole into sharp focus, adjusting the brightness as
needed to optimize the image contrast.
Figure 95.
Backlit pinhole (in focus but not centered)
9. Position the focused pinhole exactly in the cross hairs of the microscope eyepiece.
Look through the eyepiece on the microscope and adjust the diopter ring on the right lens
if necessary to straighten the cross hairs.
Adjust the X and Y position of the stage until the pinhole is
exactly
under the box in the
center of the cross hairs.
Figure 96.
Pinhole in focus and in position
Pinhole
Cross hairs
Pinhole
Содержание DXR xi
Страница 12: ...1 Introducing the DXRxi Raman Imaging System Trademarks 6 DXRxi User Guide Thermo Scientific...
Страница 14: ...2 DXRxi Microscope Operation 8 DXRxi User Guide Thermo Scientific...
Страница 64: ...Safety Information 58 DXRxi User Guide Thermo Scientific Microscope Safety Labels or...
Страница 65: ...Safety Information Thermo Scientific DXRxi User Guide 59...
Страница 66: ...Safety Information 60 DXRxi User Guide Thermo Scientific...
Страница 67: ...Safety Information Thermo Scientific DXRxi User Guide 61...
Страница 68: ...Safety Information 62 DXRxi User Guide Thermo Scientific...
Страница 166: ...3 OMNICxi Software Operation Acquire and Analyze Data 160 DXRxi User Guide Thermo Scientific...