3
OMNICxi Software Operation
Verify Instrument Performance
Thermo Scientific
DXRxi User Guide
165
Click the PV status icon to access features for
and
To run performance verification
1. Set up the microscope:
• Make sure the correct laser, filter and grating are installed in the instrument
• Set up the microscope for reflected light brightfield illumination (see the
DXRxi
Quick Reference Guide
for details)
• Set the
to its middle position
• Make sure the
is installed on the microscope stage
2. Click the PV status icon and choose
Run
.
A message similar to the following is displayed:
Table 39.
PV status, available options
Icon
Meaning
PV information for current laser set is valid (all tests passed) and up-to-date
PV information for current laser set is invalid (one or more tests failed during
last PV run)
PV information for current laser set is either out-of-date or this laser set has
never been validated
No PV tests available for current laser set
Note
: Only full range gratings can be verified. PV is also unavailable for the
fiber probe.
Note
The Run button is disabled if
PV status icon
Содержание DXR xi
Страница 12: ...1 Introducing the DXRxi Raman Imaging System Trademarks 6 DXRxi User Guide Thermo Scientific...
Страница 14: ...2 DXRxi Microscope Operation 8 DXRxi User Guide Thermo Scientific...
Страница 64: ...Safety Information 58 DXRxi User Guide Thermo Scientific Microscope Safety Labels or...
Страница 65: ...Safety Information Thermo Scientific DXRxi User Guide 59...
Страница 66: ...Safety Information 60 DXRxi User Guide Thermo Scientific...
Страница 67: ...Safety Information Thermo Scientific DXRxi User Guide 61...
Страница 68: ...Safety Information 62 DXRxi User Guide Thermo Scientific...
Страница 166: ...3 OMNICxi Software Operation Acquire and Analyze Data 160 DXRxi User Guide Thermo Scientific...