System Tools
162
DXRxi User Guide
Thermo Scientific
6. From the OMNICxi Acquisition window, click
Settings
and then click
.
The Setup Video window is displayed with an area for the video image and features for
calibrating it.
Figure 100.
Setup Video window
7. In the Setup Video window, set
Current calibrations
to the correct calibration for the
10x objective.
8. Position the calibration slide under the white light from the illuminator.
Turn up the brightness to optimize the contrast. When you look at the calibration slide
directly (not through the eyepiece), you should see a pool of white light from the
illuminator on the surface of the slide.
Continue looking at the calibration slide directly while using the joystick and focus
control on the stage controller to roughly center the large circle in the middle of the slide
under the white light.
Note
If you are calibrating the 10x objective for the first time, skip this step.
Tip
Hold down the button to the left of the joystick or focus control to speed up the
respective control.
Содержание DXR xi
Страница 12: ...1 Introducing the DXRxi Raman Imaging System Trademarks 6 DXRxi User Guide Thermo Scientific...
Страница 14: ...2 DXRxi Microscope Operation 8 DXRxi User Guide Thermo Scientific...
Страница 64: ...Safety Information 58 DXRxi User Guide Thermo Scientific Microscope Safety Labels or...
Страница 65: ...Safety Information Thermo Scientific DXRxi User Guide 59...
Страница 66: ...Safety Information 60 DXRxi User Guide Thermo Scientific...
Страница 67: ...Safety Information Thermo Scientific DXRxi User Guide 61...
Страница 68: ...Safety Information 62 DXRxi User Guide Thermo Scientific...
Страница 166: ...3 OMNICxi Software Operation Acquire and Analyze Data 160 DXRxi User Guide Thermo Scientific...