Chapter 7
Column
Packed Columns
128
Packed Columns
There are different sizes of packed columns with both metric and imperial
dimensions with dedicated adapters. The FOCUS accepts 1/4 inch OD, 1/8 inch
OD imperial metal packed columns, 6 mm OD metric metal packed columns. The
metal packed columns can be installed into the S/SL injector by using the
appropriate conversion kit.
Metric Packed Columns
The metal 6 mm OD are the metric packed columns most commonly used:
Using Correct Metric Fittings
To connect packed columns to injector and detector base bodies, you must use the
correct column ferrules and retaining nuts.
Metric Column Ferrules
Metal packed columns require double brass ferrules (front and back)
Metric Retaining Nuts
Use hexagonal 1/4 inch column retaining nuts to connect all metal packed
columns.
Table 7-2 lists the correct fittings for metric packed columns while Figure 7-2
shows the fittings.
Table 7-2. Metric Packed Column Fittings
Column Type
Ferrules
Retaining Nut
6 mm OD metal column
brass double
hexagonal 1/4 inch
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...