Chapter 9
Thermal Conductivity Detector (TCD)
TCD Menu
180
Transf temp
50–375 ºC
This is the transfer line temperature for the heated
zone between the oven and the detector cell. Set
ON
to turn on the heater and display the actual and
setpoint values. Set
OFF
to turn off the heater. A value
between the oven temperature, and the detector
temperature must be set.
The suggest value is 10 °C lower than the Block temp
setpoint.
Const fil temp?
Yes/No
Set
YES
to activate the constant filament temperature
mode and display the
Fil temp (CT)
parameter.
The filament voltage is automatically set to 15V.
Set
NO
to display the filament voltage and maximum
filament temperature parameters. The current
operating mode (
CT
,
CV
, or
CC
) is displayed in
parentheses.
Fil temp (--)
2
On/Off,
50–450 ºC
This parameter indicates the filament temperature.
Fil volts (--)
3
5–15 V
This parameter indicates the filament voltage.
Fil temp limit
3-4
50–450 ºC
This parameter indicates the maximum filament
temperature.
Ref flow
On/Off,
This parameter indicates the reference gas flow. You
can only turn the flow on and off. Set
ON
to turn on
the flow. Set
OFF
to turn off the flow.
Mkup flow
On/Off,
This parameter indicates the make-up gas flow. You
can only turn the flow on and off. Set
ON
to turn on
the flow. Set
OFF
to turn off the flow.
1.
Filament power cannot be turned
On
if either carrier gas and reference gas are not
On
.
2.
This parameter appears only if the
Const fil temp?
parameter is set to
Yes
.
3.
This line appears only if
Const fil temp?
is set to
No
.
4.
When the TCD with the polyimide coated filaments is used, the max temperature is 320 °C
Table 9-4. The Detector (TCD) Parameters (Continued)
Menu
Range
Comments
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...