Chapter 15
Maintenance and Troubleshooting
TCD Troubleshooting
300
TCD Troubleshooting
Sensitivity
Sensitivity is related to the detector temperature (increasing temperature reduce
sensitivity) and to the flow rate of the carrier, reference and make-up gas. A gain
in sensitivity can be obtained increasing the gap between the temperature of the
block and that of the filaments or increasing the filaments voltage.
Sensitivity is strictly related to the state of the filaments and to their operating
conditions. A significant reduction of the detector sensitivity may be caused by
the contamination of the filaments due to degradation of high molecular weight
compounds inside the cell or to contaminated gases.
Low temperatures of the detector block may cause high boiling compounds
condensate on the filaments reducing sensitivity.
Baseline drift
A small baseline drift normally occurs during a temperature program and does not
indicate a trouble. This effect is due to the decrease of the carrier gas flow rate as
temperature increases.
A baseline that suddenly goes out of scale and a rapid growth of the signal could
indicate that filaments are likely to be burnt and have to be replaced.
Negative peaks
Negative peaks are normally generated by the sample components that have a
thermal conductivity higher than carrier gas. For instance, using nitrogen or argon
as carrier gas, negative peaks are obtained for helium, hydrogen or methane.
To revert the polarity of the detector, refer to
Chapter 9.
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...