Chapter 15
Maintenance and Troubleshooting
Maintaining a SSL Injector
268
Removing the Septum
4. Unscrew the septum cap.
5. Remove the septum holder with septum, then the septum support. See A of
Figure 15-2.
Figure 15-2. Replace and Clean the Liner (1)
6. Use the liner cap wrench provided with the GC to remove the injector liner
cap. See B of Figure 15-2.
INLET
Temp 25 Off
CARRIER
Pressure 0 Off
B
A
C