Chapter 15
Error Messages
Maintenance and Troubleshooting
307
•
Check the column efficiency periodically.
Detector
•
Ensure that the detector parameters are correctly set.
•
Check for eventual detector contamination.
Data System
•
Check that the Data System in use is properly connected.
Instrument Troubleshooting
The control unit of the GC performs a number of test routines to monitor and
report on the status of the major chromatographic and control functions.
Instrument malfunctioning, due to a component failure or to abnormal operating
condition is identified by the red lighting of the Not Ready/Error LED located on
the GC Status Panel.
When lit, this LED indicates that the injector and oven power has been cut off for
safety reasons and an error message indicating the possible cause of error appears
on the display. See
Error Messages
The following table reports the error messages and the explanation of the relevant
correlated problem
Table 15-6. Error Messages
Message
Description
TEMP SHUTDOWN
Loss of carrier
This message appears in two cases:
The carrier supply may be missing and the actual pressure value is 0, or
the actual pressure cannot reach the setpoint value.
HARDWARE SHUTDOWN
Oven T over limits
The oven temperature exceeds the temperature limits (350 °C). Error
may be due to a probe malfunctioning.
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...