Chapter 14
SOP number: P0474/05/E - 02 May 2007
252
12. In Main Menu move the cursor until
Detector
is selected. Press
ENTER
to
open
DETECTOR
menu.
13. Scroll to
Block temp, Transf temp, Reference
,
Make-up, Const
fil temp, Fil volts (CV), Fil temp limit
and
Filament
power
. By using the
INC
and
DEC
keys select the value or the required
operating mode, then press
ENTER
.
1). In case of TCD with the polyimide coated filaments, the temperature limit is 320 °C.
FOCUS GC
Detector
DETECTOR (TCD)
Block temp
200 200
DETECTOR (TCD)
Transf temp
190 190
DETECTOR (TCD)
Reference
On
DETECTOR (TCD)
Make-up
On
DETECTOR (TCD)
Cost fil temp.
N
DETECTOR (TCD)
Fil volt (CV)
10
DETECTOR (TCD)
Fil temp limit
350
1
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...