Chapter 10
AI 3000/AS 3000 Menu
Autosamplers
193
AI 3000/AS 3000 Menu
This paragraph describes the AI 3000/AS 3000 menu.
AUTOSAMPLER
Sample volume 1.0
Sample rinses 10
Plunger strokes 10
Pre wash solv A+B
Pre wash cycles 10
Post wash solv A+B
Post wash cycles 10
Viscous sample No
Sampl.depth Bottom
Inj. Depth Standard
Pre dwell time 10
Post dwell time 10
Table 10-3. AI 3000/AS 3000 Menu
Parameter
Range
Description
AUTOSAMPLER
This line is the title bar
Sample volume
0 to 5 µl with the
10 µl syringe
0 to 2.5 µl with the
5 µl syringe
0 to 250 nl with the
0.5 µl syringe
This parameter specifies the amount of sample to be
injected for each run.
Sample rinses
0–15 times
This parameter specifies the number of times the
syringe is pre-washed with sample.
Plunger strokes
0–15 times
The number of times the syringe plunger will be
pulled up and down to eliminate any bubbles from
the sample.
Содержание FOCUS GC
Страница 1: ...Thermo Scientific FOCUS GC Gas Chromatograph Instruction Manual PN 31709412 Revision May 2007 HOME ...
Страница 4: ...iv InstructionManual ...
Страница 70: ...Chapter 2 The FOCUS GC User Interface Info Diagnostics 70 Instruction Manual ...
Страница 92: ...Chapter 4 Column Oven Oven Parameters 92 Instruction Manual ...
Страница 124: ...Chapter 6 Purged Packed Column Injector PPKD PPKD Injector Menu 124 Instruction Manual ...
Страница 170: ...Chapter 8 Flame Ionization Detector FID Detector Signal Parameters 170 Instruction Manual ...
Страница 186: ...Chapter 9 Thermal Conductivity Detector TCD Detector Signal Parameters 186 Instruction Manual ...
Страница 242: ...Chapter 13 SOP number P0473 04 E 02 May 2007 242 Instruction Manual ...
Страница 320: ...Chapter 15 Maintenance and Troubleshooting Guide to Solve Analytical Problems 320 Instruction Manual ...
Страница 336: ...Chapter 16 LAN Option Set up Set Up 336 Instruction Manual ...
Страница 344: ...Glossary 344 Instruction Manual C min 1 instead of C min g L 1 instead of g L ...
Страница 350: ...Index 350 Instruction Manual ...